USER MANUAL
PORTABLE
DISSOLVED OXYGEN METER
MODEL: MW600
Smart DO Meter
Przeglądanie stron 0
1 2 3 4 5 6 ... 11 12

Podsumowanie treści

Strona 1 - Milwaukee

MilwaukeeUSER MANUALPORTABLEDISSOLVED OXYGEN METERMODEL: MW600Smart DO Meter

Strona 2 - PROBE PREPARATION:

BATTERY REPLACEMENT:When the battery becomes weak themeter will display the low batteryindicator " ".When this appears, only a few hoursof b

Strona 3 - CALIBRATION PROCEDURE:

SPECIFICATIONS:MW600RANGE 0.0 to 19.9 mg/LRESOLUTION 0.1 mg/LACCURACY (@25°C) ±1.5% Full ScaleTEMPERATURE Automatic from 0 to 30°CCOMPENSATIONCALIBRAT

Strona 4

ISTMW600 01/10MilwaukeeAuthorized Dealer:

Strona 5

PROBE PREPARATION:The meter is supplied with a 9V battery. Slide off the batterycompartment cover on the back of the meter. Install the batteryinto th

Strona 6 - ALTITUDE COMPENSATION:

PROBE POLARIZATION:• A dissolved oxygen probe must be polarized to functionproperly.• To polarize the probe, the fully assambled probe must beconnecte

Strona 7 - SALINITY COMPENSATION:

Zero Calibration:• Dip the probe into MA9070 zero oxygensolution and stir gently for 2-3 minutes.• Wait for the probe to reach it’s loweststable readi

Strona 8

TAKING MEASUREMENTS:Verify the probe is polarized and the probe andmeter have been calibrated. Remove the protectivecap from probe. Immerse the tip of

Strona 9

5°C02468101214161820222426283032343638400 m14.613.813.112.411.811.310.810.39.99.59.18.78.48.17.87.57.37.16.86.66.4300 m14.113.312.712.011.410.910.49.9

Strona 10

6SALINITY COMPENSATION:The table below illustrates the change in the solubility of oxygenin air saturated water as a result of chloride concentration

Strona 11

7PROBE & MEMBRANE MAINTENANCE:The D.O. probe body is made of reinforced plastic for maximumdurability.A thermistor temperature sensor provides tem

Strona 12

The Platinum cathodeshould always be brightand untarnished. If it istarnished or stained, dueto contact with certaingases or a damagedmembrane cap, th